Z-1035-2016 Class I Terminated

Recalled by Verathon, Inc. — Bothell, WA

FDA device recall Z-1035-2016 was initiated by Verathon, Inc. on Jan 29, 2016 and is designated Class I. Reason for recall: The firm has become aware of the potential for disruption (flickering) in the video laryngoscopy image when GlideScope Titanium SU laryngoscope blades are in use. Video "flickering" appears as the… The recall status is terminated (terminated Mar 1, 2017). Affected quantity: 7733 units (6377 units in the US and 1356 units International).

Recall details

Product Type
Devices
Report Date
Mar 16, 2016
Initiation Date
Jan 29, 2016
Termination Date
Mar 1, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7733 units (6377 units in the US and 1356 units International)

Product description

GlideScope Titanium Single-Use Video Laryngoscope. It is a single-use video laryngoscope system (GlideScope Titanium System) consists of a sterile-packaged, single-use video laryngoscope, multi-use Smart Cable and portable GlideScope video monitor. The system incorporates a high-resolution, full-color digital camera and monitor for real-time viewing and recording. Affected Product Names and Part Numbers: -LoPro S3: 0574-0130 (single blade) or 0270-0769 (box of 10) -LoPro S4: 0574-0131 (single blade) or 0270-0770 (box of 10) -MAC S3: 0574-0132 (single blade) or 0270-0771 (box of 10) -MAC S4: 0574-0133 (single blade) or 0270-0772 (box of 10) The GlideScope Video Laryngoscopes are intended for use by qualified medical professionals to obtain a clear, unobstructed view of the vocal cords for medical procedures.

Reason for recall

The firm has become aware of the potential for disruption (flickering) in the video laryngoscopy image when GlideScope Titanium SU laryngoscope blades are in use. Video "flickering" appears as the intermittent break-up of the on-screen video image, appearing as either distorted horizontal or vertical bars of displaced video signal.

Distribution pattern

Worldwide Distribution - US (nationwide) and to countries of: Australia, Belgium, Canada, Denmark, France, Germany, Hong Kong, Italy, Luxembourg, Saudi Arabia, Singapore, Spain, Sweden, The Netherlands, U.A.E., and United Kingdom.

Code information

LOT Number ranges: -LoPro S3: 081814 - 093015; -LoPro S4: 081114 - 090315; -MAC S3: 080814 - 101315; -MAC S4: 022514 - 082115;

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1035-2016
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1035-2016. Report a data issue: hello@regdatalab.com.