Z-2476-2020 Class I Terminated

Recalled by Verathon, Inc. — Bothell, WA

FDA device recall Z-2476-2020 was initiated by Verathon, Inc. on Jun 5, 2020 and is designated Class I. Reason for recall: When video laryngoscopy system users apply, twisting motions while connecting the blade in the HDMI port, or excessive torsion to the connection port, this force is transferred to the HDMI PCB, which… The recall status is terminated (terminated Apr 22, 2024). Affected quantity: 437.

Recall details

Product Type
Devices
Report Date
Jul 15, 2020
Initiation Date
Jun 5, 2020
Termination Date
Apr 22, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
437

Product description

GlideScope Core OneTouch Smart Cable, REF: 0800 0601, an accessory to the GlideScope Video Laryngoscope System Per PLRA: The Core OneTouch Cables (Reusable Device) is ancillary equipment used with the Core 10 and Core 15 monitors. The monitor and workstation are intended to work with video endoscopes, in conjunction with ancillary equipment, for endoscopic procedures.

Reason for recall

When video laryngoscopy system users apply, twisting motions while connecting the blade in the HDMI port, or excessive torsion to the connection port, this force is transferred to the HDMI PCB, which may result in video synchronization interruption, loss of live image/blank screen, no cameras connected monitor warning, may present risk to patients, in which time to intubation is more critical.

Distribution pattern

Worldwide Distribution: US (nationwide) and countries of: Australia and United Kingdom.

Code information

All serial numbers

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2476-2020
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2476-2020. Report a data issue: hello@regdatalab.com.