Z-1039-2024 Class II Ongoing
FDA device recall Z-1039-2024 was initiated by Becton Dickinson & Co. on Jan 8, 2024 and is designated Class II. Reason for recall: There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional… The recall status is ongoing. Affected quantity: 2,363,168 total units.
Recall Details
- Product Type
- Devices
- Report Date
- Feb 14, 2024
- Initiation Date
- Jan 8, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2,363,168 total units
Product Description
BD BBL Sensi Disc Sulfamethoxazole with Trimethoprim 23.75/1.25 ug, used for semi-quantitative in vitro susceptibility testing. Catalog Number 231536 and 231539¿
Reason for Recall
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
Distribution Pattern
Domestic distribution nationwide. International distribution worldwide.
Code Information
Catalog No. 231536 UDI-DI 00382902315360 Lots 1060689 1116935 1210905 1242939 1306103 1334461 2004540 2025116 2056456 2108378 2146453 2301307 3003869 3062089 3094151 3118079 3152968 3248444 0309595 0343394; Catalog No. 231539¿ UDI-DI 30382902315392 Lots 2339393 3003869 3062089 3094151 3118079 3152968 3214768 3248444 3282335 0309595 0343394 1060689 1116935 1148543 1210905 1242939 1306103 1334461 2004540 2025116 2056456 2108378 2146453 2277321 2301307
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1039-2024
- openFDA file date
- Oct 26, 2026
- Loaded by RegDataLab
- Oct 27, 2026