Z-1045-2023 Class II Ongoing

Recalled by MEDLINE INDUSTRIES, LP - Northfield — Northfield, IL

FDA device recall Z-1045-2023 was initiated by MEDLINE INDUSTRIES, LP - Northfield on Dec 22, 2022 and is designated Class II. Reason for recall: The recall is due to observed intra-operative screw failure. When pre-drilled with ¿1.3mm drill bit for ¿2.0mm screws and ¿1.6mm drill bit for ¿2.4mm we have observed screw failure due to excessive… The recall status is ongoing. Affected quantity: 420 units.

Recall Details

Product Type
Devices
Report Date
Feb 8, 2023
Initiation Date
Dec 22, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
420 units

Product Description

MEDLINE UNITE FOOT & ANKLE Screw, Non-Locking, Mini T8 (2.0x8mm), REF MMSN2008

Reason for Recall

The recall is due to observed intra-operative screw failure. When pre-drilled with ¿1.3mm drill bit for ¿2.0mm screws and ¿1.6mm drill bit for ¿2.4mm we have observed screw failure due to excessive torque.

Distribution Pattern

US Nationwide distribution.

Code Information

GTIN 10193489120387, Lot Number 810304

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1045-2023
openFDA file date
Oct 21, 2026
Loaded by RegDataLab
Oct 23, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.