Z-1070-2019 Class II Terminated

Recalled by ICU Medical, Inc. — San Clemente, CA

Recall Details

Product Type
Devices
Report Date
April 3, 2019
Initiation Date
November 27, 2018
Termination Date
February 1, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5000

Product Description

ICU Medical, Transpac IV Monitoring Kit With Safeset Reservoir and Blood Sampling Port, 152 cm (60") Tubing, Disposable Transducer, 03 ml Squeeze Flush Device, Macrodrip (Pole Mount), REF: 01C-42640-06, Sterile EO

Reason for Recall

Reports of pressure tubing separation between the sampling port and the high pressure tubing on certain lots of extravascular blood pressure transducers, which could lead to delay in therapy, and in rare circumstances, pressure tubing separations could lead to blood loss, air embolism, or contamination of the fluid path.

Distribution Pattern

U.S.: CA, TX, WI, NY, NC, GA, PA, and NY. Foreign: Canada

Code Information

Lot/UDI: 3514084/ (01)0 0840619 03747 5 (17) 201001 (30) 01 (10) 3514084; 3550279/ (01)0 0840619 03747 5 (17) 201001 (30) 01 (10) 3550279