Z-1074-2025 Class I Ongoing

Recalled by Getinge Usa Sales Inc — Wayne, NJ

FDA device recall Z-1074-2025 was initiated by Getinge Usa Sales Inc on Jan 15, 2025 and is designated Class I. Reason for recall: Sevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalation and/or skin exposure. Potential harms include… The recall status is ongoing. Affected quantity: 6 units.

Recall details

Product Type
Devices
Report Date
Feb 19, 2025
Initiation Date
Jan 15, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6 units

Product description

Vaporizer Sevoflurane, Quik-Fil. Model Number: 6682285. Device is a component of Flow-i C20, Flow-i C30, Flow-i C40, Flow-c, Flow-e. For use during anesthesia.

Reason for recall

Sevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalation and/or skin exposure. Potential harms include irritation of respiratory tract, which in worst case may lead to lung edema and/or severe hypocalcemia, blistering, superficial ulceration, and/or hypomagnesemia.

Distribution pattern

Domestic distribution to Kentucky and Connecticut.

Code information

Model Number: 6682285. UDI-DI: 07325710001141. Serial Numbers: 3761, 7635, 7633, 7630, 7632, 7631.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1074-2025
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Getinge Usa Sales Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Getinge Usa Sales Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Getinge Usa Sales Inc →

Source: openFDA record for Z-1074-2025. Report a data issue: hello@regdatalab.com.