Z-0458-2025 Class II Ongoing
FDA device recall Z-0458-2025 was initiated by Getinge Usa Sales Inc on Oct 3, 2024 and is designated Class II. Reason for recall: Error code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand… The recall status is ongoing. Affected quantity: 22 units.
Recall details
- Product Type
- Devices
- Report Date
- Nov 20, 2024
- Initiation Date
- Oct 3, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 22 units
Product description
Reason for recall
Distribution pattern
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bhutan, Brazil, Canada, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Ecuador, Egypt, France, Germany, Greece, India, Israel, Japan, Jordan, Korea, Republic Of Kuwait, "Macedonia, The Former Yugoslav Republic Of", Malaysia, Mauritius, Mexico, Montenegro, Netherlands, New Zealand, Panama, Peru, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Viet Nam.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0458-2025
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
Need the full FDA record for Getinge Usa Sales Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Getinge Usa Sales Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.