Z-1079-2014 Class II Terminated

Recalled by Zimmer, Inc. — Warsaw, IN

FDA device recall Z-1079-2014 was initiated by Zimmer, Inc. on Oct 17, 2013 and is designated Class II. Reason for recall: As a result of complaints for disassociation of the TM Reverse Glenosphere from the Baseplate, Zimmer updated the related surgical technique and released a video to provide additional instructions to… The recall status is terminated (terminated Sep 12, 2014). Affected quantity: 73,002 units.

Recall Details

Product Type
Devices
Report Date
Mar 5, 2014
Initiation Date
Oct 17, 2013
Termination Date
Sep 12, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
73,002 units

Product Description

Trabecular Metal" Reverse Shoulder System

Reason for Recall

As a result of complaints for disassociation of the TM Reverse Glenosphere from the Baseplate, Zimmer updated the related surgical technique and released a video to provide additional instructions to users on how to properly assemble the Glenosphere to the Baseplate.

Distribution Pattern

Worldwide Distribution-USA (nationwide) and the countries of Germany, India, Malaysia, China, Singapore, Korea, Taiwan, and Brazil.

Code Information

61025613, 61009830, 61038635, 61031385, 61021507, 61021506, 61016907, 61009834, 61006493, 61006492, 60999239, 60993837, 60993836, 60987295, 60982839, 60982838, 60970974, 60970973, 60950253, 60964425, 60946627, 60970946, 60964426, 60943695, 60940448, 60950251, 60946624, 60940449, 60927405, 60934492, 60900542, 60927406, 60919029, 60914617, 60919028, 60900543, 60893180, 60893181, 60906211, 60887745, 60887744, 60875337, 60875336, 60868040, 60873954, 60868041, 60867386, 60867385, 60856882, 60856880, 60855532, 60852815, 60855529, 60846158, 60846157, 60840619, 60834839, 60806512, 60824952, 60820211, 60814537, 60785523, 60795705, 60792083, 60761203, 60752933, 60781096, 60776583, 60765951, 60768804, 60757206, 60747093, 60741590, 60747092, 60741589, 60738183, 60727960, 60738184, 60730072, 60730073, 60723302, 60723277, 60718989, 60712568, 60683321, 60675599, 60669310, 60674468, 105533, 60605155, 105588, 60577323, 60580942, 60572796, 60564704, 60572798, 60568536, 60568535, 60578677, 60568468, 60568467, 60568466, 60564623, 60564639, 60564628, 60564641, 60564638, 60564636, 60568470, 60564642, 60564635, 60564625, 60564621, and 60564633.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1079-2014
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.