Z-1092-2014 Class II Terminated
FDA device recall Z-1092-2014 was initiated by Zimmer, Inc. on Jan 10, 2014 and is designated Class II. Reason for recall: This single manufacturing lot has the potential for the bond strength between the Trabecular Metal material and substrate to be below the minimum required specification. Testing found bond strength… The recall status is terminated (terminated Sep 11, 2014). Affected quantity: 10 units.
Recall Details
- Product Type
- Devices
- Report Date
- Mar 5, 2014
- Initiation Date
- Jan 10, 2014
- Termination Date
- Sep 11, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 10 units
Product Description
00434811113, REF I4348-111-13 Zimmer Trabecular Metal Shoulder Humeral Stem 48 Degrees 11 mm Stem Diameter 130 mm Stem Length Trabecular Metal Humeral Stem 48 Degree 11mm x 130mm, Sterile, QTY 1, Rx Product Usage: Zimmer Trabecular Metal TM Humeral Stem is a shoulder prosthetic replacement device that is indicated for the treatment of severe pain or significant disability in degenerative, rheumatoid, or traumatic disease of the glenohumeral joint as well as humeral fractures. The humeral stem uses. Trabecular Metal around the proximal part of the stem for biological ingrowth.
Reason for Recall
This single manufacturing lot has the potential for the bond strength between the Trabecular Metal material and substrate to be below the minimum required specification. Testing found bond strength approximately 80% of the minimum required by specification in one of the three zones of the Trabecular Metal. The other two zones exceeded the minimum specification.
Distribution Pattern
Distribution USA in the state of Illinois.
Code Information
lot 62267250
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1092-2014
- openFDA file date
- Oct 23, 2026
- Loaded by RegDataLab