Z-1093-2016 Class II Terminated

Recalled by Biomet, Inc. — Warsaw, IN

FDA device recall Z-1093-2016 was initiated by Biomet, Inc. on Feb 16, 2016 and is designated Class II. Reason for recall: Zimmer Biomet Inc. Verso Shoulder Forked Retractor (Comprehensive Instrumentation 2 Prong Retractor) supplied by Kirkstall Precision was made according to the W1.4034 (420 stainless) which does not… The recall status is terminated (terminated May 30, 2017). Affected quantity: 1261 units.

Recall Details

Product Type
Devices
Report Date
Mar 16, 2016
Initiation Date
Feb 16, 2016
Termination Date
May 30, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1261 units

Product Description

Verso Shoulder Forked Retractor (Comprehensive Instrumentation 2 Prong Retractor) Biomet Comprehensive Reverse Shoulder products are indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The patient must be anatomically and structurally suited to receive the implants and a functional deltoid muscle is necessary. The Comprehensive Reverse Shoulder is indicated for primary, fracture, or revision total shoulder replacement for the relief of pain and significant disability due to gross rotator cuff deficiency.

Reason for Recall

Zimmer Biomet Inc. Verso Shoulder Forked Retractor (Comprehensive Instrumentation 2 Prong Retractor) supplied by Kirkstall Precision was made according to the W1.4034 (420 stainless) which does not conform to 420 S29 indicated on the print. All units manufactured to date were manufactured using 420 stainless. Therefore, all lots of PN: 402852 are considered in scope.

Distribution Pattern

Worldwide Distribution - US including CA, NY, VA, IL, IN, SC, NC, AR, MO, ND, AL, PA, MI, WA, OR, LA, FL, GA, OH, KS, KY, TX, MA, AZ, NJ, UT, IO, NM, WV, MS, NE, and MD OUS: Australia, Chile, Columbia, Costa Rica, Argentina, Korea, Japan, Panama, Malaysia, Puerto Rico, Canada, and The Netherlands.

Code Information

all lots of PN: 402852

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1093-2016
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.