Z-1099-2017 Class II Terminated

Recalled by Roche Diabetes Care, Inc. — Indianapolis, IN

FDA device recall Z-1099-2017 was initiated by Roche Diabetes Care, Inc. on Dec 20, 2016 and is designated Class II. Reason for recall: iOS and Android: Under certain conditions the affected app versions may disregard historical bolus data potentially leading to an incorrect bolus insulin recommendation being provided to the user… The recall status is terminated (terminated Sep 5, 2017). Affected quantity: 27243.

Recall Details

Product Type
Devices
Report Date
Feb 1, 2017
Initiation Date
Dec 20, 2016
Termination Date
Sep 5, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
27243

Product Description

Accu-Chek Connect Diabetes Management App

Reason for Recall

iOS and Android: Under certain conditions the affected app versions may disregard historical bolus data potentially leading to an incorrect bolus insulin recommendation being provided to the user. iOS only: Pairing and using multiple meters with the Accu-Chek Connect app can under rare circumstances cause the bolus advisor to fail to offer a correction bolus recommendation within the eligible time window following a blood glucose measurement (10  15 minutes). Depending on the individual metabolic situation potentially incorrect bolus advice could lead to serious health consequences such as hypoglycemia. Both software issues may also cause the amount of active insulin displayed during the bolus calculation process to be incorrect and should not be used to manually calculate a bolus.

Distribution Pattern

Worldwide Distribution - Nationwide Distribution The Accu-Chek Connect App for iOS is distributed directly to end users via the Apple App Store. The Accu-Chek Connect App for Android is distributed directly to end users via the Google Play App Store.

Code Information

Catalog number 07562462001 / GTIN number 00365702700000 Catalog number 07250452001 / GTIN number 00365702700017

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1099-2017
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.