Z-1125-2017 Class II Terminated

Recalled by Covidien LLC — North Haven, CT

FDA device recall Z-1125-2017 was initiated by Covidien LLC on Dec 16, 2016 and is designated Class II. Reason for recall: Labeling error: The label indicates the product is latex free, but the headband component included with the forehead sensor does contain latex. The recall status is terminated (terminated Dec 4, 2018). Affected quantity: 7453 Units.

Recall Details

Product Type
Devices
Report Date
Feb 8, 2017
Initiation Date
Dec 16, 2016
Termination Date
Dec 4, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7453 Units

Product Description

Covidien Nellcor SpO2 Forehead Sensor Item Code: RS10; The Covidien Nellcor¿ Forehead SpO2Sensor, model RS10, is indicated for use when continuous non-invasive arterial oxygen saturation and pulse rate monitoring are required for patients weighing more than 40 kg.

Reason for Recall

Labeling error: The label indicates the product is latex free, but the headband component included with the forehead sensor does contain latex.

Distribution Pattern

Nationwide. Foreign: Belgium Germany France Netherlands

Code Information

Lot Numbers: 161510095H, 161620089H, 161690203H, 161760132H, 161900091H,161970090H, 162040107H, 162110103H, 162320191H, 162390212H