Z-1126-2019 Class II Terminated
FDA device recall Z-1126-2019 was initiated by Ulrich Medical USA Inc on Mar 8, 2019 and is designated Class II. Reason for recall: Certain Torque Limiting Handles in the field are beyond their 3 year calibration life. The recall status is terminated (terminated Jul 10, 2020). Affected quantity: 25.
Recall details
- Product Type
- Devices
- Report Date
- Apr 17, 2019
- Initiation Date
- Mar 8, 2019
- Termination Date
- Jul 10, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 25
Product description
ULRICH Torque Limiting Handle for use with a Set Screw Driver to perform the final tightening of set screws in a spinal implant construct.
Reason for recall
Certain Torque Limiting Handles in the field are beyond their 3 year calibration life.
Distribution pattern
AZ, IL, LA, MI, MO, NV, OH, PA, TN, and TX.
Code information
Model #s CS 7130 Serial #s 538436-02_70 (69 total) UDI Description 4052536123704, Model #s CS 7134 Serial #s 538438-02_70 (69 total) UDI Description 4052536123728, Model #s CS 7146 Serial #s 04-14-022-18_23 (6 total) UDI Description4052536078301, Model #s CS 7148 Serial #s 11-12-035-01_09, 03-12-006-24, and 03-12-006-25 (11 total) UDI Description 4052536078318, and Model #s CS 8049 Serial #s J04192-31_40, J04192-056, J04192-064, H12015-031_6, and J01276-031_3, (21 total) UDI Description 4052536022076.
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1126-2019
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026
Need the full FDA record for Ulrich Medical USA Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.