Z-1350-2020 Class II Terminated

Recalled by Ulrich Medical USA Inc — Chesterfield, MO

FDA device recall Z-1350-2020 was initiated by Ulrich Medical USA Inc on Dec 31, 2019 and is designated Class II. Reason for recall: Due to a production error, the weld seam of the "Locking screw for crosslink, screw-to-screw" cannot reliably transmit the required torque of 3.5 Nm during installation. The recall status is terminated (terminated Nov 17, 2025). Affected quantity: 179 units.

Recall details

Product Type
Devices
Report Date
Mar 4, 2020
Initiation Date
Dec 31, 2019
Termination Date
Nov 17, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
179 units

Product description

Ulrich medical neon3 OCT spinal stabilization, REF CS 3917-12, Locking screw for crosslink, screw-to-screw

Reason for recall

Due to a production error, the weld seam of the "Locking screw for crosslink, screw-to-screw" cannot reliably transmit the required torque of 3.5 Nm during installation.

Distribution pattern

AZ, TX, GA, FL, AR, CA

Code information

Lot Numbers: U010577, U016620, U019084

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1350-2020
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Ulrich Medical USA Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1350-2020. Report a data issue: hello@regdatalab.com.