Z-1350-2020 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- March 4, 2020
- Initiation Date
- December 31, 2019
- Termination Date
- November 17, 2025
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 179 units
Product Description
Ulrich medical neon3 OCT spinal stabilization, REF CS 3917-12, Locking screw for crosslink, screw-to-screw
Reason for Recall
Due to a production error, the weld seam of the "Locking screw for crosslink, screw-to-screw" cannot reliably transmit the required torque of 3.5 Nm during installation.
Distribution Pattern
AZ, TX, GA, FL, AR, CA
Code Information
Lot Numbers: U010577, U016620, U019084