Z-1148-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- March 28, 2018
- Initiation Date
- January 12, 2018
- Termination Date
- August 28, 2018
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 16
Product Description
Artis one fluoroscopic x-ray system and patient table Artis one is an angiography system developed for diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Arts one include cardiac angiography, neuro-angiography, general angiography, rotational angiography; multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities.
Reason for Recall
Certain patient table systems may have a missing or incorrectly installed snap ring that could cause the bolt to slide out of its base and the tabletop carriage to slide to the floor.
Distribution Pattern
US Distribution to the states of : AL, AR, CA, FL, IN, LA, MI, MO, NC, NY, PA, TX, and WA.
Code Information
Serial numbers: 82044 82105 82070 82213 82054 82040 82038 82205 82065 82267 82108 82010 82039 82109 82102 82066