Z-1154-2024 Class II Ongoing

Recalled by Canon Medical System, USA, INC. — Tustin, CA

FDA device recall Z-1154-2024 was initiated by Canon Medical System, USA, INC. on Nov 15, 2023 and is designated Class II. Reason for recall: VAREX changed the manufacturing of the internal board, and the detectors displayed abnormal images. Internal board of the X-ray flat panel detector may display abnormal images, such as horizontal… The recall status is ongoing. Affected quantity: 1.

Recall details

Product Type
Devices
Report Date
Feb 28, 2024
Initiation Date
Nov 15, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1

Product description

The device is an X-ray system that is capable of radiographic and fluoroscopic studies and is used in an interventional setting. The system consists of a C-arm which is equipped with an X-ray tube, beam limiter and X-ray receptor, X-ray controller, computers with system and processing software, and a patient radiographic table.

Reason for recall

VAREX changed the manufacturing of the internal board, and the detectors displayed abnormal images. Internal board of the X-ray flat panel detector may display abnormal images, such as horizontal stripes, which may lead to the inability to complete an ongoing exam, or the patient may need to be examined with another x-ray system.

Distribution pattern

US: CA, LA

Code information

UDI-DI: 04987670100192, Serial Number: B1E22X2005

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1154-2024
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Canon Medical System, USA, INC.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1154-2024. Report a data issue: hello@regdatalab.com.