Z-1157-2014 Class II Terminated

Recalled by Smith & Nephew Inc. — Saint Petersburg, FL

FDA device recall Z-1157-2014 was initiated by Smith & Nephew Inc. on Jan 9, 2014 and is designated Class II. Reason for recall: Reports of wound fluid and /or blood that were not evacuated from beneath the wound dressing and the pump blockage alarm did not activate has been associated with inadequate Soft Port sealing at the… The recall status is terminated (terminated Sep 26, 2014). Affected quantity: 162,629 invididual units.

Recall Details

Product Type
Devices
Report Date
Mar 19, 2014
Initiation Date
Jan 9, 2014
Termination Date
Sep 26, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
162,629 invididual units

Product Description

Smith & Nephew RENASYS Soft Port REF: 66800799 (The RENASYS Soft Port Dressing Kits are used in conjunction with negative pressure wound therapy pumps)

Reason for Recall

Reports of wound fluid and /or blood that were not evacuated from beneath the wound dressing and the pump blockage alarm did not activate has been associated with inadequate Soft Port sealing at the wound site or improper positioning of the Soft Port orifice over the wound contact dressing drape. Inadequate sealing can result in air leaks and contribute to inefficient fluid evacuation from the wo

Distribution Pattern

Worldwide Distribution-USA (nationwide) including Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, ID, IL, KY, LA, MA, MS, MT, NC, NJ, NV, NY, OH, OR, PA, TN, TX, VA, WA, WI, WV, HI and the countries Dubai, Curacao, Argentina, Australia, Chile, Germany, France, England, Hong Kong, Kuwait, Lebanon, Libya, Sri Lanka, Malaysia, New Zealand, Manila, Saudi Arabia, South Africa, Singapore, Turkey, Taipei and Vietnam.

Code Information

ALL LOT #s 2013010311 and lower

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1157-2014
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.