Z-1157-2018 Class II Terminated

Recalled by Bovie Medical Corporation — Clearwater, FL

FDA device recall Z-1157-2018 was initiated by Bovie Medical Corporation on Feb 8, 2018 and is designated Class II. Reason for recall: Bovie Medical is issuing a correction to address incomplete insertion of the J-Plasma(R) handpiece cable plug into the generator receptacle resulting in an incomplete helium seal. The potential exists… The recall status is terminated (terminated Sep 17, 2020). Affected quantity: 12 units.

Recall Details

Product Type
Devices
Report Date
Mar 28, 2018
Initiation Date
Feb 8, 2018
Termination Date
Sep 17, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
12 units

Product Description

J-Plasma Precise(R) FLEX, Catalog Number: BVX-500BF

Reason for Recall

Bovie Medical is issuing a correction to address incomplete insertion of the J-Plasma(R) handpiece cable plug into the generator receptacle resulting in an incomplete helium seal. The potential exists for patient fluid backflow into the handpiece and the generator receptacle leading to risk of cross-contamination if the generator is reused.

Distribution Pattern

worldwide

Code Information

All lots manufactured since 01/01/2014

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1157-2018
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.