Z-1157-2024 Class II Ongoing

Recalled by MEDLINE INDUSTRIES, LP - Northfield — Northfield, IL

FDA device recall Z-1157-2024 was initiated by MEDLINE INDUSTRIES, LP - Northfield on Jan 4, 2024 and is designated Class II. Reason for recall: Packaging material was updated for the Bulb Irrigation Syringe (DYND20125), and additional testing done at Medline Corporate indicated the potential for the sterile barrier to be breached during… The recall status is ongoing. Affected quantity: 4831 units.

Recall Details

Product Type
Devices
Report Date
Sep 25, 2024
Initiation Date
Jan 4, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4831 units

Product Description

Medline Contro-Bulb Syringe, REF DYND20125; bulb irrigation syringe

Reason for Recall

Packaging material was updated for the Bulb Irrigation Syringe (DYND20125), and additional testing done at Medline Corporate indicated the potential for the sterile barrier to be breached during transportation. This only affects two lot numbers (96922110001 and 96923100001).

Distribution Pattern

US, Nationwide

Code Information

UDI/DI 40080196031765 (case), 10080196031764 (unit) Lot Numbers: 96922110001 and 96923100001

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1157-2024
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.