Z-1162-2025 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- February 19, 2025
- Initiation Date
- January 15, 2025
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 64 units
Product Description
MOSAIQ Oncology Information System
Reason for Recall
Image offsets calculated by oncology Information System software may be incorrect for Linear Accelerator (LINACs) not characterized at IEC61217 scale.
Distribution Pattern
US: PA, AL, FL, IA UT OUS: Austria Belgium Bolivia Bosnia and Herzegowina Brazil Canada France Germany Hong Kong Italy Japan Korea, Republic of Malta Netherlands Panama South Africa Spain Taiwan Thailand Turkey Venezuela
Code Information
Software versions /UDI: 3.1.2.0 (01)07340201500026 3.1.3.0 (01)07340201500026; 3.1.3.1 (01)07340201500026; 3.1.3.2 (01)07340201500026; 3.1.3.3 (01)07340201500026; 3.1.3.4 (01)07340201500026; 3.2.1.0 (01)07340201500071; 3.2.1.1 (01)07340201500071; 3.2.1.2 (01)07340201500071; 3.2.1.3 (01)07340201500071; 3.2.1.4 (01)07340201500071