Z-1167-2015 Class II Terminated

Recalled by Navilyst Medical, Inc — Glens Falls, NY

Recall Details

Product Type
Devices
Report Date
March 4, 2015
Initiation Date
January 9, 2015
Termination Date
March 10, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
64 units

Product Description

NAMIC Convenience Kit, RADIOLOGY KIT, UPN H749601014911, REF/Catalog No. 60101491, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136683. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.

Reason for Recall

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Distribution Pattern

Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.

Code Information

Lots 4268921, 4310402, 4317001, 4384580, 4442479, 4500790, 4504919, 4599070, 4645482, 4661978, 4680493, 4704381, 4742117, 4777180, 4801317.