Z-1170-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- April 4, 2018
- Initiation Date
- March 9, 2018
- Termination Date
- November 3, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 17 units
Product Description
Proclaim DRG Implantable Pulse Generator, Model Number 3664
Reason for Recall
The firm received complaints of error messages that occurred during routine impedance checks on Proclaim DRG IPGs, model 3664. Some complaints were also associated with transient over stimulation which created discomfort for the patients.
Distribution Pattern
US
Code Information
Lot Number 6210849 Serial Number AVM174.1 Lot Number 6210852 Serial Numbers: AVA560.1, AVA214.1, AVL027.1 Lot Number 6499208 Serial Numbers: AVD426.1, AVA922.1 Lot Number 6263128 Serial Numbers: AVB119.1, AVD400.1, AVA622.1, AVD431.1, AVE633.1, AVK678.1, AVN837.1 Lot Number 6268723 Serial Numbers: AUV770.1, AVW805.1 Lot Number 6278155 Serial Number AVK896.1