Z-1176-2018 Class II Terminated

Recalled by ConMed Corporation — Utica, NY

FDA device recall Z-1176-2018 was initiated by ConMed Corporation on Feb 23, 2018 and is designated Class II. Reason for recall: Wire sets of certain devices were assembled with adhesive on the contact wires, which may impede conductivity and could cause failure The recall status is terminated (terminated Sep 17, 2020). Affected quantity: 18,720 units in total.

Recall Details

Product Type
Devices
Report Date
Apr 4, 2018
Initiation Date
Feb 23, 2018
Termination Date
Sep 17, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
18,720 units in total

Product Description

MINI INFANT < 3KG, RADIOTRANSLUCENT ELECTRODE, Philips Plug Style Connector, REF/Catalog Number 2602H Indicated for use in external pacing, defibrillation and monitoring applications as a non-sterile, disposable device for single pediatric use, only. The PadPro 2602 electrodes provide the conductive interface between the defibrillator and/or the external transcutaneous (non-invasive) cardiac pacemaker and the pediatric patients skin. The electrode is intended for use on pediatric infants whose weight is less than 10 kg. (22 lbs.).

Reason for Recall

Wire sets of certain devices were assembled with adhesive on the contact wires, which may impede conductivity and could cause failure

Distribution Pattern

Worldwide Distribution: US (Nationwide) and countries of: New Zealand, Saudi Arabia, Switzerland, United Arab Emirates Uruguay & Republic of Macedonia (The former Yugoslav Republic of Macedonia).

Code Information

Lot Numbers: Y01221601, Y03241702, Y10121501, Y020713-17, Y04071501, & Y12021501.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1176-2018
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.