Z-1306-2019 Class II Terminated

Recalled by Conmed Corporation — Utica, NY

FDA device recall Z-1306-2019 was initiated by Conmed Corporation on Mar 21, 2019 and is designated Class II. Reason for recall: Voids in the seal or a partial seal results in an open channel and may compromise the sterility of the product The recall status is terminated (terminated Jan 17, 2023). Affected quantity: 5034.

Recall details

Product Type
Devices
Report Date
May 15, 2019
Initiation Date
Mar 21, 2019
Termination Date
Jan 17, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5034

Product description

ConMed Anchor Tissue Retrieval System, 15 MM, 1850 ML (3/BX) Catalog Number:TRS190SB2 The Anchor Tissue Retrieval System" by CONMED is a sterile disposable pouch used with a dedicated introducer for the encapture and removal of an organ or tissue from the body cavity during laparoscopic surgery.

Reason for recall

Voids in the seal or a partial seal results in an open channel and may compromise the sterility of the product

Distribution pattern

Worldwide Distribution: US (Nationwide) and countries of: Austria, Canada, Denmark, Great Britain, Ireland, Japan, Netherlands, OMAN, Poland, QATAR, ROMANIA, SINGAPORE, Spain and Sweden.

Code information

Lot codes for product manufactured to and including the dates listed below, with the exclusions noted below for each catalog number: Beginning Manufacture Date: April 26, 2018 Beginning Lot Code: 20180426X Ending Manufacture Date: February 15, 2019 Ending Lot Code: 20190215X Excluded Anchor Tissue Retrieval System" Lot Codes - 1. All alpha/numeric lot codes (e.g. lots 68A8T or 19A8T) 2. All numeric lot codes ending with the number 9 (e.g. Lot code 201902159) 3. All excluded numeric lot codes listed in the tables below for each specific catalog number: Excluded Lot Codes for Cat. Number TRS175SB2: Excluded Lot Codes for Cat. Number TRS190SB2 201807114, 201807275, 201807175 201807304

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1306-2019
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Conmed Corporation? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works