Z-1185-2021 Class II Terminated

Recalled by Brainlab AG — Munich

Recall Details

Product Type
Devices
Report Date
March 17, 2021
Initiation Date
March 1, 2021
Termination Date
March 20, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
73 ExacTrac Dynamic systems (versions 1.0.0, 1.0.1, 1.0.2) have been manufactured and distributed

Product Description

ExacTrac Dynamic (Medical Charged-Particle Radiation Therapy System).

Reason for Recall

In case of a failed automatic marker detection, a software error causes parts of the display to incorrectly behave as if the current patient position is within predefined tolerances and may allow the user to proceed to treatment despite potentially exceeding shift values.

Distribution Pattern

USA: (CA, NJ, AZ, AR, NY, IN, OR, NC), Australia, Austria, Belgium, Denmark, France, Germany, Hong Kong, Italy, Slovakia, Spain, Thailand, Turkey, United Kingdom, United Arab Emirates.

Code Information