Z-1185-2021 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- March 17, 2021
- Initiation Date
- March 1, 2021
- Termination Date
- March 20, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 73 ExacTrac Dynamic systems (versions 1.0.0, 1.0.1, 1.0.2) have been manufactured and distributed
Product Description
ExacTrac Dynamic (Medical Charged-Particle Radiation Therapy System).
Reason for Recall
In case of a failed automatic marker detection, a software error causes parts of the display to incorrectly behave as if the current patient position is within predefined tolerances and may allow the user to proceed to treatment despite potentially exceeding shift values.
Distribution Pattern
USA: (CA, NJ, AZ, AR, NY, IN, OR, NC), Australia, Austria, Belgium, Denmark, France, Germany, Hong Kong, Italy, Slovakia, Spain, Thailand, Turkey, United Kingdom, United Arab Emirates.
Code Information
Model/catalogue numbers: 20910-06 ETD Implanted lic. UDI: The GTIN for ExacTrac Dynamic version 1.0 is 04056481142315. GMDN code: 40887. Software revisions of affected ExacTrac Dynamic software: versions 1.0.0, 1.0.1, 1.0.2.