Z-1846-2021 Class II Terminated

Recalled by Brainlab AG — Munich, N/A

Recall Details

Product Type
Devices
Report Date
June 16, 2021
Initiation Date
May 14, 2021
Termination Date
March 20, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
22 systems in the U.S.

Product Description

ExacTrac Dynamic software, Model 20910-01B ETD Positioning and Monitoring Installer 1.0.2 - Product Usage: intended to position patients at an accurately defined point within the treatment beam of a medical accelerator for stereotactic radiosurgery or radiotherapy procedures, to monitor the patient position, and to provide a beam hold signal in case of a deviations in order to treat lesions, tumors, and conditions anywhere in the body when radiation treatment is indicated.

Reason for Recall

Display of potential patient movement might be delayed to the user for high dose treatments.

Distribution Pattern

US Nationwide distribution in the states of AZ, AR, CA, FL, IN, NJ, NY, NC, OR, PA, and TX. There was no government/military distribution.

Code Information

Version 1.02, GTIN 04056481142315