Z-1200-2018 Class II Terminated

Recalled by Siemens Medical Solutions USA, Inc. — Mountain View, CA

Recall Details

Product Type
Devices
Report Date
May 16, 2018
Initiation Date
March 5, 2018
Termination Date
February 26, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
229 systems

Product Description

ACUSON SC2000 Ultrasound System. The firm name on the label is Siemens Medical Solutions USA, Inc., Buffalo Grove, IL.

Reason for Recall

The ECG signal may flatline due to electromagnetic interference during the use of electrosurgical equipment.

Distribution Pattern

Distribution was made to medical facilities in AZ, CA, CT, FL, IL, IN, KS, KY, MI, MN, MO, NJ, NY, NC, OH, PA, TX, VA, and WA. There was no military/government distribution. Foreign distribution was made to Argentina, Australia, Austria, Belgium, Bosnia Herzeg, Colombia, Croatia, Denmark, France, Germany, India, Ireland, Israel, Italy, Japan, Mexico, New Zealand, Philippines, Poland, Republic Korea, Romania, Singapore, Slovakia, Spain, Sweden, Taiwan, Thailand, Turkey, UAE, and Vietnam.

Code Information

Software versions VB20A, VB20B, or VB20C that have a common physio module (CPM) (known by customers as the ECG (physio) hardware) installed at revision 4