Z-1201-2017 Class II Terminated

Recalled by Terumo Cardiovascular Systems Corporation — Elkton, MD

FDA device recall Z-1201-2017 was initiated by Terumo Cardiovascular Systems Corporation on Aug 20, 2014 and is designated Class II. Reason for recall: Terumo Cardiovascular Systems (Terumo CVS) is recalling certain lots of Hercules TM 360 Universal Stabilizing Arms because the two pins that hold the sleeve attached to the distal end of the device… The recall status is terminated (terminated Feb 15, 2017). Affected quantity: The total number of distributed devices stands at 260.

Recall Details

Product Type
Devices
Report Date
Feb 22, 2017
Initiation Date
Aug 20, 2014
Termination Date
Feb 15, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
The total number of distributed devices stands at 260

Product Description

Hercules 360 Universal Stabilizing Arm, top mount, 360-degree rotation, reusable stainless steel

Reason for Recall

Terumo Cardiovascular Systems (Terumo CVS) is recalling certain lots of Hercules TM 360 Universal Stabilizing Arms because the two pins that hold the sleeve attached to the distal end of the device may come loose, causing the pins and sleeve to separate from the device. While the arm can still function without the pins and sleeve in place, there is potential for the loose components to fall into the surgical site.

Distribution Pattern

Terumo Cardiovascular has identified 16 domestic consignees and 30 international consignees .

Code Information

Serial Number Range 70194 through 70504

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1201-2017
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.