Z-1208-2023 Class II Ongoing

Recalled by Coltene Whaledent Inc — Cuyahoga Falls, OH

FDA device recall Z-1208-2023 was initiated by Coltene Whaledent Inc on Jan 20, 2023 and is designated Class II. Reason for recall: Screw connection between the turbine head and the push button may loosen and fall off during the treatment, and increase the risk of swallowing or inhaling part of the turbines head or the push button The recall status is ongoing. Affected quantity: 3896 units.

Recall Details

Product Type
Devices
Report Date
Mar 8, 2023
Initiation Date
Jan 20, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3896 units

Product Description

SciCan STATIS SL S10001 Dental Handpieces-Intended for the removal of carious material, reducing hard tooth structure, cavity preparation, finishing tooth preparations, restorations and for polishing teeth. Model Number: S10001

Reason for Recall

Screw connection between the turbine head and the push button may loosen and fall off during the treatment, and increase the risk of swallowing or inhaling part of the turbines head or the push button

Distribution Pattern

Nationwide Foreign: Canada Germany Indonesia Ireland Lebanon Saudia Arabia Turkey

Code Information

UDI-DI: +ESANS100010 All Serial Numbers

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1208-2023
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 24, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.