Z-1226-2018 Class II Terminated
FDA device recall Z-1226-2018 was initiated by Zimmer Biomet, Inc. on Dec 20, 2017 and is designated Class II. Reason for recall: Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the… The recall status is terminated (terminated Apr 22, 2020). Affected quantity: 3454.
Recall Details
- Product Type
- Devices
- Report Date
- Apr 4, 2018
- Initiation Date
- Dec 20, 2017
- Termination Date
- Apr 22, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3454
Product Description
ACE Trochanteric Nail System STER TROC END CAP STER TROC END CAP 5MM STER TROC END CAP 10MM STER TROC END CAP 15MM STER TROC END CAP 20MM STER TROC END CAP 25MM The Trochanteric Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures and combinations of these fractures. The Trochanteric Long Nail System is additionally indicated to treat pertrochanteric fractures associated with shaft fractures, pathologic fractures in osteoporotic bone of the trochanteric and diaphyseal areas, long subtrochanteric fractures, ipsilateral femoral fractures, proximal or distal nonunions and malunions and revision procedures.
Reason for Recall
Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.
Distribution Pattern
USA (nationwide)
Code Information
903208000 903208010 903208015 903208020 903208025
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1226-2018
- openFDA file date
- Oct 23, 2026
- Loaded by RegDataLab