Z-1235-2017 Class II Terminated

Recalled by Terumo Cardiovascular Systems Corporation — Elkton, MD

FDA device recall Z-1235-2017 was initiated by Terumo Cardiovascular Systems Corporation on May 27, 2015 and is designated Class II. Reason for recall: Terumo Cardiovascular Systems Corporation has received complaints indicating that the Sarns Centrifugal Pump (disposable pump heads) have exhibited leaks. The leaks were detected during priming of the… The recall status is terminated (terminated Feb 23, 2017). Affected quantity: 579 units have entered into the distribution chain.

Recall Details

Product Type
Devices
Report Date
Mar 1, 2017
Initiation Date
May 27, 2015
Termination Date
Feb 23, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
579 units have entered into the distribution chain

Product Description

The Sarns Centrifugal Pump is a single use device sterilized by ethylene oxide gas. The device is sold single sterile in cartons of 8 units, or may be provided in bulk, non-sterile to another Terumo location for inclusion in Cardiovascular Procedure Kits. Cardiovascular Procedure kits are packaged individually.

Reason for Recall

Terumo Cardiovascular Systems Corporation has received complaints indicating that the Sarns Centrifugal Pump (disposable pump heads) have exhibited leaks. The leaks were detected during priming of the bypass circuit and during CPB procedures. In each of the reported incidents, there were no reports of patient injury as a result of the centrifugal pump leaks. Terumo Cardiovascular System's investigation into the incident has identified the cause of the leaks to be cracks in both the Top Housing and Magnet Back Housing of the pump. When the cracks cause the housing to become compromised, it is possible for fluid to leak from the pump head.

Distribution Pattern

Terumo CVS has identified 33 domestic consignees and 0 international consignee that are affected by this recall

Code Information

TAOS, TC03

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1235-2017
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.