Z-1245-2013 Class I Terminated

Recalled by Abbott Diabetes Care, Inc. — Alameda, CA

FDA device recall Z-1245-2013 was initiated by Abbott Diabetes Care, Inc. on Apr 15, 2013 and is designated Class I. Reason for recall: AT rare, extremely high glucose levels (1024 mg/dL and above), the FreeStyle InsuLinx Blood Glucose Meter may provide an inaccurate reading. The recall status is terminated (terminated Aug 3, 2015). Affected quantity: 335,273 meters; 209,336 remain at firm.

Recall details

Product Type
Devices
Report Date
May 15, 2013
Initiation Date
Apr 15, 2013
Termination Date
Aug 3, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
335,273 meters; 209,336 remain at firm.

Product description

FreeStyle InsuLinx Blood Glucose Monitoring System: For In Vitro Diagnostic Use only. Use only with FreeStyle InsuLinx test strips. Abbott Diabetes Care, Inc. 1360 South Loop Road Alameda, CA 94502 USA The FreeStyle InsuLinx Blood Glucose Test Strips are for use with the FreeStyle InsuLinx Blood Glucose Meter to quantitatively measure glucose in capillary whole blood samples drawn from the fingertip.

Reason for recall

AT rare, extremely high glucose levels (1024 mg/dL and above), the FreeStyle InsuLinx Blood Glucose Meter may provide an inaccurate reading.

Distribution pattern

Nationwide Distribution

Code information

Part Numbers: 71142-70 71143-70 71145-70 71150-70: All serial and lot numbers.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1245-2013
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Abbott Diabetes Care, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1245-2013. Report a data issue: hello@regdatalab.com.