Z-1274-2023 Class I Ongoing

Recalled by Abbott Diabetes Care, Inc. — Alameda, CA

FDA device recall Z-1274-2023 was initiated by Abbott Diabetes Care, Inc. on Feb 13, 2023 and is designated Class I. Reason for recall: Lithium-ion batteries in glucose monitoring system readers may swell, overheat, or pose fire hazard. Only use the charging cable/power adapter included with readers. Do not use reader if it is… The recall status is ongoing. Affected quantity: 1,373,311.

Recall details

Product Type
Devices
Report Date
Apr 12, 2023
Initiation Date
Feb 13, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,373,311

Product description

FreeStyle Libre Reader, REF: 71936-01, 71937-01, 71938-01 a component of the FreeStyle Libre 14 Day Flash Glucose Monitoring System

Reason for recall

Lithium-ion batteries in glucose monitoring system readers may swell, overheat, or pose fire hazard. Only use the charging cable/power adapter included with readers. Do not use reader if it is damaged, cracked, swelling, too hot to hold, it will no longer hold charge, or it fails Reader Test. Do not use other chargers. User Manual update includes instructions on reader safe storage, charging, use.

Distribution pattern

U.S. Nationwide.

Code information

REF/UDI-DI/Lot: 71936-01/357599000981/01C083D, 02C054D, 03C033D, 03B356D, 01B320D, 03B266D, 02B250D, 02B181D, 01B161D, 01B096D, 05B081D, 03B035D, 02A314D, 02A238D, 02A191D, 01A160D, 03A132D, 01A093D, 02A085D, 01A074D, 02A064D, 02A063D, 02A016D, 04Z351D, 01Z318D, 01Z282D, 01Z246D, 04Z191D, 03Z178D, 03Z151D, 03Z128D, 05Z115D, 03Z057D; 71937-01/357599000974/02B099D, 01B082D, 02A098D, 03A097D, 02A094D, 02A092D, 03A093D, 04A091D, 01A091D, 02A056D, 01A055D, 05A050D, 01A049D, 01Z360D, 01Z352D, 01Z346D, 02Z346D, 01Z344D, 02Z339D, 01Z338D, 03Z337D, 03Z336D, 03Z318D, 01Z311D, 03Z309D, 01Z305D, 05Z304D, 03Z303D, 02Z290D, 03Z290D, 02Z289D, 03Z246D, 01Z238D, 02Z178D, 03Z177D, 02Z177D, 03Z172D, 05Z172D, 02Z163D, 02Z151D, 03Z116D, 01Z116D, 01Z102D, 04Z083D, 02Z057D, 01Z056D, 02Z036D, 01Z024D, 01Y861D, 02Y854D, 03Y845D, 02Y845D, 01Y845D, 03Y839D, 01Y838D, 06Y830D, 03Y820D, 01Y820D, 02Y811D, 03Y811D, 01Y811D, 06Y795D, 05Y795D, 02Y791D, 03Y790D, 02Y789D, 01Y782D, 02Y782D, 01Y781D, 01Y776D, 01Y763D, 02Y762D, 02Y763D; 71938-01/357599002008/04C216D, 02C189D, 04C160D, 01C116D, 04C059D, 01C045D, 01C040D, 01C041D, 01C038D, 02C040D, 01C037D, 02C028D, 02C025D, 02C020D, 02C021D, 02C014D, 03C007D, 02B344D, 03B348D, 01B347D, 01B344D, 03B341D, 01B324D, 01B322D, 03B321D, 02B317D, 02B308D, 04B308D, 01B307D, 03B289D, 01B287D, 01B286D, 02B280D, 01B265D, 01B256D, 02B254D, 01B252D, 02B252D, 02B242D, 03B237D, 01B235D, 01B218D, 01B216D, 04B211D, 01B209D, 02B202D, 01B202D, 03B194D, 01B194D, 02B190D, 03B190D, 01B189D, 01B190D, 02B189D, 05B187D, 02B187D, 01B166D, 01B167D, 02B167D, 01B165D, 01B162D, 03B160D, 01B159D, 03B158D, 01B152D, 02B152D, 02B148D, 01B148D, 01B146D, 02B146D, 01B145D, 04B134D, 01B134D, 01B130D, 03B127D, 07B120D, 02B106D, 02B105D, 03B105D, 05B099D, 03B090D, 04B090D, 01B081D, 04B076D, 02B070D, 02B074D, 02B068D, 02B067D, 02B065D, 01B063D, 02B062D, 03B062D, 04B040D, 01B040D, 01B034D, 01B035D, 03B034D, 02B027D, 02B026D, 01B015D, 04B015D, 01B013D, 03B012D, 04B012D, 01B012D, 03B009D, 01A366D, 01A351D, 01A347D, 02A339D, 01A338D, 01A337D, 01A329D, 01A324D, 02A324D, 02A317D, 03A317D, 04A302D, 01A279D, 03A279D, 06A277D, 01A262D, 02A261D, 01A248D, 01A247D, 03A244D, 04A244D, 01A240D, 04A238D, 01A238D, 04A232D, 01A232D, 03A224D, 04A197D, 04A196D, 01A196D, 01A197D, 03A191D, 01A188D, 01A178D, 01A181D, 02A181D, 03A181D, 02A176D, 05A176D, 01A175D, 02A175D, 03A176D, 04A176D, 05A168D, 02A164D, 02A141D, 02A142D, 04A140D, 04A142D, 02A140D, 01A137D, 01A136D, 02A136D, 01A135D, 01A134D, 01A132D, 02A133D, 01A133D, 06A132D, 01A125D, 02A125D, 04A122D, 03A122D, 01A121D, 02A120D, 01A120D, 02A118D, 02A119D, 01A116D, 02A115D, 01A104D, 01A100D, 04A099D, 02A099D, 01A094D, 01A097D, 02A097D, 01A092D, 03A092D, 04A090D, 05A090D, 03A079D, 04A079D, 03A077D, 04A077D, 01A076D, 03A072D, 04A064D, 01A064D, 05A059D, 01A059D, 04A058D, 05A058D, 01A057D, 02A057D, 03A056D, 03A050D, 03A036D, 01A034D, 02A034D, 02A029D, 01A030D, 02A030D, 01A028D, 01A029D, 02A028D, 03A030D, 01A015D, 01A013D, 01A014D, 03A010D, 02A008D, 01A009D, 01A008D, 02A007D, 01A003D, 07Z351D, 01Z345D, 03Z345D, 01Z339D, 02Z338D, 02Z331D, 01Z331D, 03Z330D, 04Z330D, 05Z330D, 04Z326D, 02Z326D, 02Z325D, 01Z326D, 01Z325D, 03Z319D, 05Z319D, 04Z319D, 02Z303D, 03Z281D, 02Z275D, 01Z266D, 01Z262D, 02Z261D, 03Z261D, 02Z259D, 01Z252D, 01Z241D, 03Z241D, 04Z228D, 02Z227D, 02Z226D, 01Z224D, 02Z221D, 02Z220D, 01Z217D, 04Z199D, 04Z197D, 01Z197D, 02Z191D, 02Z189D, 01Z184D, 02Z179D, 04Z179D, 01Z175D, 01Z170D, 02Z171D, 04Z165D, 05Z165D, 01Z165D, 04Z158D, 02Z155D, 01Z150D, 01Z142D, 03Z141D, 01Z138D, 04Z137D, 01Z136D, 01Z135D, 01Z133D, 01Z128D, 01Z126D, 02Z121D, 01Z112D, 03Z109D, 01Z109D, 01Z108D, 02Z107D, 02Z099D, 01Z098D, 03Z095D, 03Z094D, 01Z094D, 02Z092D, 02Z093D, 01Z087D, 03Z086D, 04Z086D, 02Z085D, 01Z083D, 03Z083D, 02Z082D, 03Z081D, 01Z081D, 02Z080D, 01Z078D, 02Z073D, 03Z072D, 03Z073D, 03Z070D, 02Z067D, 04Z066D, 05Z066D, 02Z060D, 03Z060D, 02Z059D, 01Z058D, 04Z057D, 03Z053D, 01Z050D, 02Z050D, 04Z045D, 01Z044D, 01Z045D, 02Z045D, 01Z043D, 02Z044D, 01Z042D, 02Z039D, 03Z039D, 01Z039D, 01Z038D, 02Z037D, 01Z035D, 01Z031D, 02Z030D, 01Z030D, 03Z029D, 01Z025D, 02Z025D, 04Z024D, 02Z024D, 03Z024D, 01Z023D, 02Z022D, 03Z022D, 02Z018D, 01Z021D, 02Z016D, 03Z017D, 01Z016D, 03Z015D, 01Z008D, 01Z009D, 02Z008D, 02Z007D, 02Z002D, 03Z002D, 01Z002D, 01Y865D, 02Y865D, 03Y862D, 02Y860D, 01Y860D, 02Y855D, 05Y854D, 01Y854D, 01Y855D, 04Y854D, 04Y851D, 05Y851D, 02Y848D, 01Y847D, 01Y834D, 03Y834D, 02Y833D, 01Y833D, 05Y832D, 02Y831D, 01Y831D, 03Y819D, 04Y819D, 02Y810D, 03Y809D, 04Y809D, 01Y803D, 02Y803D, 03Y802D, 04Y802D, 01Y788D, 02Y788D, 01Y785D, 01Y777D, 02Y776D, 01Y775D, 01Y774D, 01Y761D, 02Y761D, 03Y760D

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1274-2023
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Abbott Diabetes Care, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Abbott Diabetes Care, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Abbott Diabetes Care, Inc. →

Source: openFDA record for Z-1274-2023. Report a data issue: hello@regdatalab.com.