Z-1246-2022 Class II Ongoing

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

Recall Details

Product Type
Devices
Report Date
June 15, 2022
Initiation Date
April 11, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
13 units

Product Description

Artis icono floor-angiography systems developed for single and biplane diagnostic imaging and interventional procedures SMN: 11327700

Reason for Recall

if the protective earth is interrupted and an additional fault leads to a fault current, the detection of the fault current may not be ensured may cause an electric shock if a person is in contact with metallic parts of the Carm

Distribution Pattern

US Nationwide distribution.

Code Information

UDI-DI: 04056869149325 Serial Number: 170026 170305 170307 170309 170310 170023 170304 170025 170051 170312 170027 170303 170053