Z-1246-2022 Class II Ongoing

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

FDA device recall Z-1246-2022 was initiated by Siemens Medical Solutions USA, Inc on Apr 11, 2022 and is designated Class II. Reason for recall: if the protective earth is interrupted and an additional fault leads to a fault current, the detection of the fault current may not be ensured may cause an electric shock if a person is in contact… The recall status is ongoing. Affected quantity: 13 units.

Recall Details

Product Type
Devices
Report Date
Jun 15, 2022
Initiation Date
Apr 11, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
13 units

Product Description

Artis icono floor-angiography systems developed for single and biplane diagnostic imaging and interventional procedures SMN: 11327700

Reason for Recall

if the protective earth is interrupted and an additional fault leads to a fault current, the detection of the fault current may not be ensured may cause an electric shock if a person is in contact with metallic parts of the Carm

Distribution Pattern

US Nationwide distribution.

Code Information

UDI-DI: 04056869149325 Serial Number: 170026 170305 170307 170309 170310 170023 170304 170025 170051 170312 170027 170303 170053

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1246-2022
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.