Z-1249-2024 Class II Ongoing

Recalled by Canon Medical System, USA, INC. — Tustin, CA

FDA device recall Z-1249-2024 was initiated by Canon Medical System, USA, INC. on Jan 29, 2024 and is designated Class II. Reason for recall: For some MRI systems, it has been found that some of the maximum Spatial Field Gradient (SFG) values displayed in "System Information" of the operation window and some of the maximum SFG values… The recall status is ongoing. Affected quantity: 1.

Recall Details

Product Type
Devices
Report Date
Mar 13, 2024
Initiation Date
Jan 29, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1

Product Description

MRI system: Vantage Elan, Model: MRT-2020

Reason for Recall

For some MRI systems, it has been found that some of the maximum Spatial Field Gradient (SFG) values displayed in "System Information" of the operation window and some of the maximum SFG values described in the safety manual are lower than the correct values, which may cause a patient with an MR conditional device to feel some discomfort during the MR scan.

Distribution Pattern

Worldwide - US Nationwide distribution including in the states of MD, IL, WV, CA, AZ, PR, CO, GA, PA, WA, UT, OK, KY, IA, IN, TN, CT, WI, OH, AL, MT, NM, TX, MO, MN, OR, ID, WY, NV, AR, FL, SD, MI, NY, NJ, KS, LA, NH, VT, MS, SC, ME and the country of DO.

Code Information

UDI-DI: 04987670101632. Serial Number: S1F2052753

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1249-2024
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.