Z-1255-2021 Class II Terminated

Recalled by R & D Systems, Inc. — Minneapolis, MN

FDA device recall Z-1255-2021 was initiated by R & D Systems, Inc. on Jan 20, 2021 and is designated Class II. Reason for recall: R&D Systems, Inc. received two complaints that the standard signal was low and controls were not within the specification range provided with the kit. Investigation confirmed low standard signal and… The recall status is terminated (terminated Feb 13, 2023). Affected quantity: 156 kits.

Recall details

Product Type
Devices
Report Date
Mar 24, 2021
Initiation Date
Jan 20, 2021
Termination Date
Feb 13, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
156 kits

Product description

Quantikine IVD Human sTfR Immunoassay (Human sTransferrin ELISA Kit (sTfR immunological test system)). An enzyme linked immunosorbent assay (ELISA) for the quantitative determination of Soluble Transferrin Receptor (sTfR) concentration in human serum and plasma as an aid in the diagnosis of anemia and polycythemia.

Reason for recall

R&D Systems, Inc. received two complaints that the standard signal was low and controls were not within the specification range provided with the kit. Investigation confirmed low standard signal and controls running out of specification for kits returned from the customer and the retained kits.

Distribution pattern

USA: (FL, LA, CA, OH), CHINA, UNITED KINGDOM, JAPAN, SINGAPORE, Jakarta Selatan,

Code information

Catalog #: DTFR1; Unique Device Identification (UDI) #: 815762021657; Lot #: P252478; P263156; P256349; P263156; P263160; P264604

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1255-2021
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for R & D Systems, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1255-2021. Report a data issue: hello@regdatalab.com.