Z-1255-2021 Class II Terminated

Recalled by R & D Systems, Inc. — Minneapolis, MN

Recall Details

Product Type
Devices
Report Date
March 24, 2021
Initiation Date
January 20, 2021
Termination Date
February 13, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
156 kits

Product Description

Quantikine IVD Human sTfR Immunoassay (Human sTransferrin ELISA Kit (sTfR immunological test system)). An enzyme linked immunosorbent assay (ELISA) for the quantitative determination of Soluble Transferrin Receptor (sTfR) concentration in human serum and plasma as an aid in the diagnosis of anemia and polycythemia.

Reason for Recall

R&D Systems, Inc. received two complaints that the standard signal was low and controls were not within the specification range provided with the kit. Investigation confirmed low standard signal and controls running out of specification for kits returned from the customer and the retained kits.

Distribution Pattern

USA: (FL, LA, CA, OH), CHINA, UNITED KINGDOM, JAPAN, SINGAPORE, Jakarta Selatan,

Code Information

Catalog #: DTFR1; Unique Device Identification (UDI) #: 815762021657; Lot #: P252478; P263156; P256349; P263156; P263160; P264604