Z-1390-2022 Class III Ongoing

Recalled by R & D Systems, Inc. — Minneapolis, MN

Recall Details

Product Type
Devices
Report Date
July 27, 2022
Initiation Date
June 14, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
400

Product Description

Quantikine IVD Erythropoietin Human Serum Controls for use as quantitative controls for the determination of Erythropoietin concentrations in human serum and plasma. IFU, Part Number 750095

Reason for Recall

The Instructions for Use (IFU, Part Number 750095) provided with the affected lots by the firm contains an error associated with product s CE marking. The CE mark contained in the IFU is followed by the Notified Body designation, British Standards Institution number 2797 (BSI-2797) for product inserts 750095.28 and 750095.29. The erroneous label CE2797 were in the IFU insert revision 750095.28 for the 408 affected lots that were shipped. No product was packed with erroneous insert revision 750095.29. Since the product is currently CE marked as a Class I - self-certified product under the EU In Vitro Diagnostics Directive, the Notified Body designation should not be specified. As part of the corrective action, the firm has notified the customers who received the products with erroneous labeling, and is conducting the replacements with the correct product insert containing the Insert Revision 750095.30

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the states of FL, MN, UT, OH, MA, TX, PR, PA, NJ, CA, WI, LA, Puerto Rico, and the countries of CANADA, UNITED KINGDOM, ISRAEL, MEXICO, THAILAND, CHINA, HONG KONG, AUSTRALIA, INDONESIA.

Code Information

Lot Code: UDI 00815762021671, Lot Codes: P301801; P304262; P306132; P310038; P314685; P320216; P322531 Expiration Date: 08/04/2023