Z-1260-2019 Class II Terminated

Recalled by GE Healthcare, LLC — Waukesha, WI

FDA device recall Z-1260-2019 was initiated by GE Healthcare, LLC on March 20, 2019 and is designated Class II. Reason for recall: During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector moun… The recall status is terminated (terminated October 5, 2020). Affected quantity: 3 units.

Recall Details

Product Type
Devices
Report Date
May 8, 2019
Initiation Date
March 20, 2019
Termination Date
October 5, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3 units

Product Description

Discovery NM/CT 670 ES Product Usage: The GE Discovery NM/CT 670 system is a medical tool intended for use by appropriately trained health professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy. The GE Discovery NM/CT 670 system, combining Nuclear Medicine (NM) and Computed Tomography (CT) systems, is intended to produce: The GE Discovery NM/CT 670 system is a medical tool intended for use by appropriately trained health professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy. The GE Discovery NM/CT 670 system, combining Nuclear Medicine (NM) and Computed Tomography (CT) systems, is intended to produce:

Reason for Recall

During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.

Distribution Pattern

US and Austria, Canada, Croatia, France, Germany, India, Italy, Japan, Republic of Korea, Netherlands, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK

Code Information

Mfg. Lot or Serial # System ID Model, Catalogue or Code # ESDX35020 403777D670C H3100BL ESPX54001 A51773134 H3100BM ESPY54002 DE203284 H3100BM