Z-1261-2015 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- March 18, 2015
- Initiation Date
- February 12, 2015
- Termination Date
- February 11, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 409
Product Description
Autostainer Link 48 (AS480), Autostainer Plus Link Instrument (AS100), Autostainer (S3400), and Autostainer Plus (S3800). Automated slide stainer for in vitro diagnostic use.
Reason for Recall
A false negative result affecting the diagnosis may occur. A defect in the syringe assembly stopcock component may, in some circumstances, cause leakage of excess buffer onto slide location 34 on the AS480, or slide locations 35 and 36 on the AS100, S3400, and S3800. Affected dates are from 12/2013-08/2014.
Distribution Pattern
Worldwide Distribution: US (nationwide) and countries of: AT, AU, BE, CA, CH, DE, DK, ES, FR, GB, IE, IT, JP, NL, NO, PL, and SE.
Code Information
992640 Rev. 02, 992085 Rev.no H, 992625 Rev.no F, 992326 Rev.no C Affected dates: 12/2013-08/2014 Each device is labeled with a unique serial number. Catalog/Model Number: - AS480 - S3800 - S3400 - AS100