Z-0287-2013 Class II Terminated

Recalled by Dako North America Inc. — Carpinteria, CA

FDA device recall Z-0287-2013 was initiated by Dako North America Inc. on May 1, 2008 and is designated Class II. Reason for recall: Dako is recalling the PT Link Waterbath because of a possible electrical hazard. Specifically, if the heater pad under the reagent tank becomes wet, it may present an electrical shock hazard when… The recall status is terminated (terminated Nov 13, 2012). Affected quantity: 136 units.

Recall details

Product Type
Devices
Report Date
Nov 21, 2012
Initiation Date
May 1, 2008
Termination Date
Nov 13, 2012
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
136 units

Product description

PT Link Waterbath, Model#PT100. Lab Vision PT Module Plus provides an automated standarized alternative to labor intensive and time consuming dewaxing and epitote retrieval procedures.

Reason for recall

Dako is recalling the PT Link Waterbath because of a possible electrical hazard. Specifically, if the heater pad under the reagent tank becomes wet, it may present an electrical shock hazard when the lid is open.

Distribution pattern

Nationwide distribution: USA including states of: AK, CA, CO, CT, DC, FL, GA, HI, IA, IL IN, KS, KY, MA, MD, MI, MN, MO, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, and WA.

Code information

Catalog #: PT-Module

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0287-2013
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Dako North America Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Dako North America Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Dako North America Inc. →

Source: openFDA record for Z-0287-2013. Report a data issue: hello@regdatalab.com.