Z-1264-2023 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- March 29, 2023
- Initiation Date
- February 15, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 7 units
Product Description
Attune Revision Limb Preservation System (LPS) Tibial Insert XSM 12MM, Part Number 151760212
Reason for Recall
The subject product lots are being recalled because they received a higher than specified irradiation dose. This exceeds the validated range for exposure to gamma radiation of these devices and may result in changes to the implant material properties.
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Canada, Poland, Germany, India, Australia, & New Zealand
Code Information
GTIN: 10603295490968; Lot: JP9016