Z-1029-2022 Class II Terminated
FDA device recall Z-1029-2022 was initiated by DePuy Orthopaedics, Inc. on Mar 11, 2022 and is designated Class II. Reason for recall: System software v1.5.1 has a system software issue related to the Daylight Savings Time (DST) change that can cause a system error, requiring the user to restart the system and potentially cause a… The recall status is terminated (terminated Apr 18, 2024). Affected quantity: US: 68 units; OUS: 10 units.
Recall details
- Product Type
- Devices
- Report Date
- May 11, 2022
- Initiation Date
- Mar 11, 2022
- Termination Date
- Apr 18, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- US: 68 units; OUS: 10 units
Product description
Reason for recall
Distribution pattern
Worldwide distribution - US Nationwide and the countries of Israel, Australia & New Zealand.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1029-2022
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026
Need the full FDA record for DePuy Orthopaedics, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.