Z-1271-2024 Class II Ongoing

Recalled by FUJIFILM Healthcare Americas Corporation — Lexington, MA

Recall Details

Product Type
Devices
Report Date
March 13, 2024
Initiation Date
January 10, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
35 Units (28 US, 7 OUS)

Product Description

Synapse PACS - Version 7.2.100

Reason for Recall

Measurements on a Secondary Capture 2D image, that does not have pixel spacing in the DICOM header, when combined with a Breast Tomo Series results in incorrect measurements.

Distribution Pattern

Worldwide distribution - US Nationwide, including Puerto Rico and the countries of Australia, Belgium, Brazil, Canada, Colombia, Ecuador, France, Germany, India, Italy, Mexico, Namibia, Netherlands, Portugal, Qatar, Spain, Thailand, Turkey, UAE, United Kingdom.

Code Information

Software Version: 7.2.100, UDI: (01)854904006008(10)0702100