Z-1299-2026 Class II Ongoing
Recalled by FUJIFILM Healthcare Americas Corporation
— Lexington, MA
FDA device recall Z-1299-2026 was initiated by FUJIFILM Healthcare Americas Corporation on Jan 9, 2026 and is designated Class II. Reason for recall: It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power… The recall status is ongoing. Affected quantity: 16 units (38 additional units 11/04/2026).
Recall details
- Product Type
- Devices
- Report Date
- Feb 11, 2026
- Initiation Date
- Jan 9, 2026
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 16 units (38 additional units 11/04/2026)
Product description
FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)
Reason for recall
It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even if the power of the device is cut off. As a result, it cannot be denied that the circuit in question overheats and damages the surrounding parts. It may also cause smoke.
Distribution pattern
US Nationwide distribution in the states of AK, CA, CT, FL, IL, NH, OH, WI.
Code information
Model Number: CH-200; (1) Version: 566-16130-23; UDI-DI: 04540217052226; Serial numbers: MP95A9F6A001, MP95A9F6B001, MP95A9F74001, MP95A9F74002, MP95A9F76001, MP95A9F78001; (2) Version: 566-16130-31; UDI-DI: 04540217057436; Serial numbers: MP95AA783001; (3) Version: 566-16130-33; UDI-DI: 04540217057450; Serial numbers: MP95AA983001, MP95AA983002, MP95AA983003, MP95AA985001, MP95AA985003, MP95AA985002, MP95AA988001, MP95AA988002, ***Serial numbers added 11/04/2026: MP95AC7D3002, MP95AC7B9003, MP95AC7D3001, MP95AC7D6001, MP95AA9A2001, MP95AC7C7001, MP95AC7D2001, MP95AA9A1001, MP95AA9A1002, MP95AC7CC003, MP95AC7D7001, MP95AC7D9001, MP95AA98A001, MP95AA993002, MP95AA996002, MP95AA996001, MP95AC7B8001, MP95AA993001, MP95AA997001, MP95AA9A2002, MP95AC7D4001, MP95AA8A3001, MP95AA8A3002, MP95AC7B9001, MP95AC7B9004, MP95AC7CB001, MP95AC7CB002, MP95AC7CB003, MP95AC7CC001, MP95AC7CC002, MP95AC7D4002, MP95AA89C001, MP95AA8A7001, MP95AC5B4001, MP95AC5B4002, MP95AC7B6001, MP95AC7B6002, MP95ABBB7007;
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1299-2026
- openFDA file date
- Nov 24, 2026
- Loaded by RegDataLab
- Nov 25, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for FUJIFILM Healthcare Americas Corporation? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.
Get a free API key
How the connector works