Z-1299-2026 Class II Ongoing

Recalled by FUJIFILM Healthcare Americas Corporation — Lexington, MA

FDA device recall Z-1299-2026 was initiated by FUJIFILM Healthcare Americas Corporation on Jan 9, 2026 and is designated Class II. Reason for recall: It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power… The recall status is ongoing. Affected quantity: 16 units (38 additional units 11/04/2026).

Recall details

Product Type
Devices
Report Date
Feb 11, 2026
Initiation Date
Jan 9, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
16 units (38 additional units 11/04/2026)

Product description

FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)

Reason for recall

It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even if the power of the device is cut off. As a result, it cannot be denied that the circuit in question overheats and damages the surrounding parts. It may also cause smoke.

Distribution pattern

US Nationwide distribution in the states of AK, CA, CT, FL, IL, NH, OH, WI.

Code information

Model Number: CH-200; (1) Version: 566-16130-23; UDI-DI: 04540217052226; Serial numbers: MP95A9F6A001, MP95A9F6B001, MP95A9F74001, MP95A9F74002, MP95A9F76001, MP95A9F78001; (2) Version: 566-16130-31; UDI-DI: 04540217057436; Serial numbers: MP95AA783001; (3) Version: 566-16130-33; UDI-DI: 04540217057450; Serial numbers: MP95AA983001, MP95AA983002, MP95AA983003, MP95AA985001, MP95AA985003, MP95AA985002, MP95AA988001, MP95AA988002, ***Serial numbers added 11/04/2026: MP95AC7D3002, MP95AC7B9003, MP95AC7D3001, MP95AC7D6001, MP95AA9A2001, MP95AC7C7001, MP95AC7D2001, MP95AA9A1001, MP95AA9A1002, MP95AC7CC003, MP95AC7D7001, MP95AC7D9001, MP95AA98A001, MP95AA993002, MP95AA996002, MP95AA996001, MP95AC7B8001, MP95AA993001, MP95AA997001, MP95AA9A2002, MP95AC7D4001, MP95AA8A3001, MP95AA8A3002, MP95AC7B9001, MP95AC7B9004, MP95AC7CB001, MP95AC7CB002, MP95AC7CB003, MP95AC7CC001, MP95AC7CC002, MP95AC7D4002, MP95AA89C001, MP95AA8A7001, MP95AC5B4001, MP95AC5B4002, MP95AC7B6001, MP95AC7B6002, MP95ABBB7007;

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1299-2026
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for FUJIFILM Healthcare Americas Corporation? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1299-2026. Report a data issue: hello@regdatalab.com.