Z-1281-2019 Class II Terminated

Recalled by Neusoft Medical Systems Co., Ltd. — Shenyang, N/A

FDA device recall Z-1281-2019 was initiated by Neusoft Medical Systems Co., Ltd. on Oct 22, 2018 and is designated Class II. Reason for recall: Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan, there will be some electromagnetic interference in the system which may affect the… The recall status is terminated (terminated Aug 25, 2022). Affected quantity: 2.

Recall details

Product Type
Devices
Report Date
May 29, 2019
Initiation Date
Oct 22, 2018
Termination Date
Aug 25, 2022
Voluntary/Mandated
FDA Mandated
Product Quantity
2

Product description

NeuViz 16 Essence Multi-slice CT Scanner System Product Usage: The Multi-Slice CT Scanner System can be used as a whole body computed tomography x-ray system featuring a continuously rotating x-ray tube and detector array. The acquired x-ray transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.

Reason for recall

Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan, there will be some electromagnetic interference in the system which may affect the control signal of the flying focal spot, resulting in an incorrect flying focal spot sequence. If the system does not receive error, it would continue to complete the scan with the incorrect flying focal spot sequence, having an effect on the reconstruction of some images, resulting in image loss.

Distribution pattern

US Nationwide Distribution.

Code information

Series numbers N16E180037E; N16E180047E

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1281-2019
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Neusoft Medical Systems Co., Ltd.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Neusoft Medical Systems Co., Ltd.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Neusoft Medical Systems Co., Ltd. →

Source: openFDA record for Z-1281-2019. Report a data issue: hello@regdatalab.com.