Z-1281-2019 Class II Terminated

Recalled by Neusoft Medical Systems Co., Ltd. — Shenyang, N/A

Recall Details

Product Type
Devices
Report Date
May 29, 2019
Initiation Date
October 22, 2018
Termination Date
August 25, 2022
Voluntary/Mandated
FDA Mandated
Product Quantity
2

Product Description

NeuViz 16 Essence Multi-slice CT Scanner System Product Usage: The Multi-Slice CT Scanner System can be used as a whole body computed tomography x-ray system featuring a continuously rotating x-ray tube and detector array. The acquired x-ray transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.

Reason for Recall

Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan, there will be some electromagnetic interference in the system which may affect the control signal of the flying focal spot, resulting in an incorrect flying focal spot sequence. If the system does not receive error, it would continue to complete the scan with the incorrect flying focal spot sequence, having an effect on the reconstruction of some images, resulting in image loss.

Distribution Pattern

US Nationwide Distribution.

Code Information

Series numbers N16E180037E; N16E180047E