Z-1650-2016 Class II Terminated

Recalled by Neusoft Medical Systems Co., Ltd. — Shenyang, N/A

Recall Details

Product Type
Devices
Report Date
May 18, 2016
Initiation Date
May 2, 2015
Termination Date
November 4, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7 units

Product Description

NeuViz 64 Multi-Slice CT Scanner System (consist if two variants: NeuViz 64e, NeuViz 64i)

Reason for Recall

It was found by R&D in April, 2014 that there was a defect in NeuViz 64 system software version 1.0.5+P09. The icons of side decubitus didn't meet the patient position description, when the scanner position was set to "Right" in the system setting - scanner options.

Distribution Pattern

US Distribution - Including Puerto Rico and the states of IL, CT, SC, NE

Code Information

Serial Numbers: N64E120022E, N64E130039E, N64I130051E, N64E130062E, N64E130063E, N64E130066E, N64I140077E