Z-1282-2019 Class II Terminated

Recalled by Medtronic Inc. — Northridge, CA

Recall Details

Product Type
Devices
Report Date
May 15, 2019
Initiation Date
July 4, 2018
Termination Date
June 6, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
Approximately 9,328 registered users

Product Description

Guardian Connect App CSS7200 used on the iPhone, iPad, and iPod Touch devices.

Reason for Recall

The application may be closed by the operating system without alerting the user the app is no longer running or communicating with the transmitter resulting in the user not receiving alerts that could be associated with hypoglycemic or hyperglycemic events.

Distribution Pattern

Distribution was only made OUS. Countries involved are: Austria, United Arab Emirate, Australia, Belgium, Switzerland, Chile, Colombia, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Hungary, Ireland, Israel, Iceland, Italy, Republic of Korea, Luxembourg, Netherlands, Norway, Poland, Qatar, Romania, Russian Federation, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Turkey, and South Africa.

Code Information

UDI - (01)00643169693142(10)