Z-1289-2024 Class II Ongoing

Recalled by Baxter Healthcare Corporation — Deerfield, IL

FDA device recall Z-1289-2024 was initiated by Baxter Healthcare Corporation on Feb 14, 2024 and is designated Class II. Reason for recall: Three lots of the Boot Assembly, Traction Device are being recalled due to due to a customer report of an incorrectly oriented locking pin which caused the boot to detach from the traction system. If… The recall status is ongoing. Affected quantity: 14 units.

Recall Details

Product Type
Devices
Report Date
Mar 20, 2024
Initiation Date
Feb 14, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
14 units

Product Description

STERIS Boot Assembly, Traction Device, OT 1000, Part Number REF 10009192

Reason for Recall

Three lots of the Boot Assembly, Traction Device are being recalled due to due to a customer report of an incorrectly oriented locking pin which caused the boot to detach from the traction system. If an impacted device is used during surgery, the boot may detach from the traction system, potentially leading to patient harm and a delay in surgery.

Distribution Pattern

US Nationwide distribution in the state of Alabama.

Code Information

UDI/DI 00615521000851, Lot Numbers: 23-140, 23-268, 23-314

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1289-2024
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.