Z-1290-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- April 11, 2018
- Initiation Date
- January 22, 2018
- Termination Date
- September 24, 2019
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 9621
Product Description
DS Titanium Ligation Clip, used in laparoscopic surgery
Reason for Recall
Added a contraindication to IFU: Do not use the DS clips for living donor nephrectomies. Added other contraindications. A publication from 2017 recommended Aesculap DS Clips for kidney transplantation with living donors. The information from this publication was not approved or validated by the firm.
Distribution Pattern
US and Canada distribution
Code Information
DS Clips - Product Numbers PL450SU DS SINGLE FIRE LIG.CLIPS PL471SU DS SINGLE FIRE LIG.CLIP L W/LATCH PL453SU DS SINGLE FIRE LIG.CLIP SM PL475SU DS SINGLE FIRE LIG.CLIP XL W/LATCH PL459SU DS SINGLE FIRE LIG.CLIP M PL462SU DS SINGLE FIRE LIG.CLIP ML PL465SU DS SINGLE FIRE LIG.CLIP ML W/LATCH PL468SU DS SINGLE FIRE LIG.CLIP L (all clips going back 5 years)