Z-1297-2022 Class II Ongoing

Recalled by Atrium Medical Corporation — Merrimack, NH

FDA device recall Z-1297-2022 was initiated by Atrium Medical Corporation on Apr 21, 2022 and is designated Class II. Reason for recall: Mismatch between the expiration date on the sterile barrier pouch and the expiration date on the case/shipper label. Product sterile barrier pouch label expiration date of 16SEP2024 (2024-09-16) while… The recall status is ongoing. Affected quantity: 56 cases (560 units / 10 units per case).

Recall Details

Product Type
Devices
Report Date
Jul 6, 2022
Initiation Date
Apr 21, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
56 cases (560 units / 10 units per case)

Product Description

Atrium Firm PVC Thoracic Catheter 28French-Intended to facilitate the evacuation of air and/or fluid from the chest cavity or mediastinum. Product REF Number: 15028

Reason for Recall

Mismatch between the expiration date on the sterile barrier pouch and the expiration date on the case/shipper label. Product sterile barrier pouch label expiration date of 16SEP2024 (2024-09-16) while the case and shipper labels both showed 21SEP2024 (2024-09-21), causing product being used for up to five (5) days past the true expiration date

Distribution Pattern

AZ IL KS LA MI NY OH OK TX VA

Code Information

UDI-DI: 20650862150289 - case / 00650862150285- each Lot Number: ME231010

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1297-2022
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.